DENSITY METER
The new Cannon CT-600 has many of the engineering and control features of the CT-500 bath, but the CT-600 also includes a built-in thermoelectric cooling unit. The CT-600 controls temperature per ASTM D 445 precision specifications between 10°C and 100°C, and is ideal for controlled temperature applications between 10°C and 30°C. When tests at low temperatures are to be made, the cooling unit is engaged via a switch on the front panel. When working at higher temperatures (30°C and above), the switch is turned off. Nothing could be simpler.
A special selector switch on the front panel facilitates rapid setting of ten popular temperatures for kinematic viscosity determinations: 10, 15, 20, 25, 30, 40, 50, 60, 80, and 100°C. After the switch has been set the bath will equilibrate to within a fraction of a degree of the desired temperature. A fine-tuning control then allows temperature adjustment to within ± 0.01°C of the target temperature as measured by the user’s reference thermometer. By switching to the “variable” setting on the front panel, any temperature within the operating range can be obtained.
Features:
Ideal for kinematic viscosity determinations between 10°C and 30°C
Built-in thermoelectric cooling provides “rock-solid” temperature control at ambient and below
Meets ASTM D 445 precision specifications from 10°C to 100°C
CT-600
Dimensions (W x D x H) 407 mm x 362 mm x 610 mm (16 in x 14.25 in x 24 in)
Weight 22 kg (48 lb) includes bath jar
Shipping Weight 34 kg (75 lb) includes bath jar
Bath Capacity 17 L (4.5 gal)
Bath Temperature Range (Stability) 10 °C to 100 °C (± 0.01 °C)
Operating Conditions 15 °C to 30 °C, 10% to 90% relative humidity (non-condensing), Installation Category II; Pollution Degree 2
Electrical Specifications 115/230 Vac, 50/60 Hz, 1000 watts
Compliance CE Mark; EMC directive (89/336/EEC); Low voltage directive (73/23/EEC); ROHS